Cirrus-logic AN365 Manual do Utilizador Página 8

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 14
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 7
AN365
8 AN365REV1
4.2 Load Performance with Rogowski Coil TR9L
4.2.1 Active Energy Load Performance
Power Factor Load Current (A) Current Dynamic Range (x:1) Error
PF = 1
120 1 0.01%
60 2 0.04%
12 10 0.06%
1.2 100 0.02%
0.24 500 0.01%
0.12 1000 0.00%
0.06 2000 -0.02%
0.036 3333 0.00%
0.03 4000 0.02%
Lagging PF = 0.5
120 1 0.05%
60 2 0.04%
12 10 0.03%
1.2 100 -0.06%
0.24 500 0.05%
0.12 1000 0.01%
0.06 2000 -0.01%
Leading PF = 0.5
120 1 0.01%
60 2 -0.02%
12 10 -0.03%
1.2 100 -0.03%
0.06 2000 -0.09%
Table 6. Active Energy Load Performance
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.01 0.1 1 10 100
Percentage Error (%)
Load Current (A)
PF = 1 Lagging PF = 0.5 Leading PF = 0.5
Figure 7. Active Energy Load Performance
Vista de página 7
1 2 3 4 5 6 7 8 9 10 11 12 13 14

Comentários a estes Manuais

Sem comentários